| Surgery Section - Wireless Pressure Monitoring
in Endovascular Aneurysm
| Topic: Wireless Pressure
Monitoring in Endovascular Aneurysm |
Date of Origin: 08/2008 |
| Section: Surgery |
Policy No: 163 |
| Approved Date: 09/16/2008 |
Effective Date: 11/01/2008 |
| Next Review Date: 09/2009 |
|
| |
IMPORTANT REMINDER
This Medical Policy has been developed through consideration of medical necessity,
generally accepted standards of medical practice, and review of medical literature
and government approval status.
Benefit determinations should be based in all cases on
the applicable contract language. To the extent there are any conflicts
between these guidelines and the contract language, the contract language will
control.
The purpose of medical policy is to provide a guide to coverage. Medical Policy
is not intended to dictate to providers how to practice medicine. Providers
are expected to exercise their medical judgment in providing the most appropriate
care.
Description
The goal of abdominal aortic aneurysm (AAA) repair is to reduce pressure
in the aneurysm sac and thus prevent rupture. Failure to completely
exclude the aneurysm from the systemic circulation results in continued
pressurization. An endoleak (persistent perfusion of the aneurysmal
sac) may be primary (within the first 30 days) or secondary (after 30
days). Endoleaks are reported to vary from 10%–50% of cases,
and there are 5 types of endoleaks. (2) Type I endoleaks result
from ineffective fixation at either end of the graft; while these can
seal spontaneously, risk of rupture is high and intervention is often
indicated. Type II endoleaks result from retrograde filling of
the aneurysm mainly from lumbar and/or inferior mesenteric arteries. Risk
of rupture is less than with types I and III and type II endoleaks can
often be monitored when the aneurysm is shrinking. Type III endoleaks
are caused by failure of the implanted graft and include development
of holes, which need to be treated aggressively. Type IV endoleaks
are caused by the porosity of the graft fabric. Type V endoleaks
are referred to as endotension and correspond to continued aneurysm expansion
in the absence of a confirmed endoleak. Endoleaks, particularly
types I and III, lead to continued sac pressurization and therefore may
be considered technical failures of endovascular aneurysm repair (EVAR).
The completeness of exclusion or absence of endoleaks is evaluated by
intraoperative angiography. However, interpretation
of images can be problematic and it can also cause patient morbidity
due to the dye load from repeated injections of contrast material. Direct
measurement of sac pressure provides a physiologic assessment of success. Studies
have used direct sac pressure measurements with a catheter; the drawback
of this approach is the interference by the catheter during endovascular
repair and the inability to leave it in place. Since endoleaks
may also develop subsequent to the time of surgery, computed tomography
(CT), magnetic resonance imaging (MRI), and ultrasound are used in monitoring
the aneurysmal sac. Percutaneous catheter-based approaches can
also be used to measure intrasac pressures postoperatively.
Several factors determine aneurysm sac pressure after EVAR. These
include graft-related factors such as endoleak, graft porosity, and graft
compliance and anatomic factors such as patency of aneurysm side branches,
aneurysm morphology, and the characteristics of aneurysm thrombus.
Given this situation, wireless implantable pressure-sensing devices
are being evaluated to monitor pressure in the aneurysm sac. These
implanted devices use various mechanisms to wirelessly transmit pressure
readings to devices for measuring and recording pressure. These
devices have the potential to improve outcomes for patients who have
had endovascular repair. They may change the need for or the frequency
of monitoring of the aneurysm sac using contrast-enhanced CT scans. They
may improve postoperative monitoring. However, the accuracy of
these devices must be determined, and potential benefits and risks must
be considered and evaluated. At present, two types of systems are
being evaluated: radiofrequency and ultrasound-based systems.
In October 2006, the U.S. Food and Drug Administration (FDA) cleared
the CardioMEMS EndoSure™ (radiofrequency based) system through
the 510(k) process. (3) The favorable FDA review indicated only
that the device was substantially equivalent to legally marketed predicate
devices. The FDA labeling indications noted that the device is
intended for measuring intrasac pressure during endovascular abdominal
aortic aneurysm repair. It also noted that it may be used as an
adjunctive tool in the detection of intraoperative endoleaks. The
ImPressure™ system (ultrasound-based) is used in Europe and is
being used as part of an investigation device exemption (IDE) trial of
stent grafts.
Policy/Criteria
Use of wireless pressure sensors is considered investigational in the
management (intraoperative and/or postoperative) of patients having endovascular
aneurysm repair.
Position Summary
While multiple factors (see above) influence aneurysm sac pressure after
endovascular aneurysm repair (EVAR), models have demonstrated that with
exclusion of the sac after EVAR, sac pressures diminish significantly. Using
direct translumbar puncture in ten patients after EVAR with sac shrinkage
and no evidence of endoleak, Sonesson demonstrated that mean intrasac
pressure diminishes to 20% of mean arterial pressure. (4) Dias
and colleagues reported on 46 percutaneous intrasac pressure readings
in 37 patients following EVAR. (5) In these patients, they calculated
the mean pressure index (MPI)—the percentage of mean intra-aneurysm
pressure relative to the simultaneous mean intra-aortic pressure. Median
MPI was 19% in 11 patients with shrinking sacs, 30% in ten patients
with unchanged sacs, and 59% in 9 expanding aneurysms. Type II
endoleaks (six patients) were associated with a wide range of
MPI (22%–92%). The authors comment that the findings from
this small series do not imply that imaging follow-up can be replaced
by pressure measurements. They also note that a definitive pressure
threshold using direct measurement for subsequent intervention needs
to be defined by further studies.
For the wireless devices, Ohki and colleagues reported initial results
from the APEX study—acute pressure measurement to confirm aneurysm
sac exclusion. (6) They reported 30-day results on 76 of 90 enrolled
patients at 12 sites worldwide who received the CardioMEMS wireless pressure
sensor during EVAR. Of the patients enrolled, results were not
reported on 14 patients due to “protocol deviations, typically
a missed measurement.” In one patient, the device could
not be deployed because there was inadequate space (less than 10 mm)
within the aneurysm sac after graft deployment. In all 76 patients,
there was close agreement between the wireless sensor and angiographic
catheter for a type I endoleak equivalent. As defined in the study,
a reduction in pulse pressure of 30% or more from the initial pressure
measurement would be associated with a sealed sac and a less than 30%
reduction in pulse pressure would indicate a type I or III endoleak. With
angiography as the standard, for detection of type I or III endoleak
at the completion of the procedure, the sensor detected four of
five (80%) of the leaks. The authors comment that the case
that was not detected was not clinically significant, i.e., no intervention
was required. The wireless sensor indicated no endoleak in 66 of
71 (93%) cases without an angiographic endoleak. Deployment of
the device was noted to add 10 minutes to the EVAR procedure; the average
operative time for those in the study was 205 +/- 87 minutes. No
complications were felt due to the wireless sensor, although the authors
did comment that there was a learning curve associated with deployment
and using the device. This report also notes that these patients
will be followed up for five years, and that long-term data will
provide information to evaluate the value of the sensor for postoperative
follow-up surveillance. The value of the device will need to consider
not only benefit but also any potential complications due to the implanted
device. A clinical trial is being initiated to assess the safety
and efficacy of the Endosure device in comparison to CT angiography for
long-term follow-up after EVAR.
Ellozy reported results with a mean follow-up of 11 months of 21 patients
using the Impressure AAA Sac Pressure Transducer. (7) This device
was studied as part of an IDE examining use of an endovascular stent-graft
in the repair of infrarenal AAAs in high-risk patients. This transducer
is hand-sewn into the outside of the stent-graft and then packaged as
part of the delivery sheath. During follow-up, pressures could
be obtained at all visits in 15 of the 21 patients. There were
problems with readings from four of the devices that were thought
to be due to placement of the devices between the iliac limbs of the
stent graft. For the 14 patients with follow-up of at least six months,
aneurysm sac shrinkage of more than five mm was seen in seven patients,
and the mean pressure index (MPI) was significantly lower in those with
sac shrinkage at six months. Two of the patients with shrinking
aneurysms had type II endoleaks.
In 2008, two case series were published, both using the EndoSure radiofrequency
device, one of intraoperative use and one of postoperative follow-up
for 30 days. The intraoperative series reported the correlation
of measurements made during the procedure using the pressure sensor and
a catheter inserted into the aneurysm sac among 19 patients. (8) Although
the authors reported that all correlation coefficients were statistically
significant, they ranged from 0.50 to 0.96. Data presented in
the paper show marked differences in measurement, suggesting that the
accuracy of the measures requires further study. Of eight sets
of measurements, four had more than 50% of patients with at least 10%
variation between methods. A second series was an Ultrasound-based
study of postoperative monitoring for endoleaks using the EndoSure sensor
in 12 patients with 30-day follow-up. (9) At 30 days, 2 type-II
endoleaks were noted on CT. Sac pressures were unchanged with
1 patient and had decreased in the other. One patient with a type-III
endoleak on CT had increasing sac pressure. Delivery of the sensor
was complicated in two of 12 patients (17%).
Additional data are needed from larger, randomized trials with long-term
follow-up to determine the effectiveness of wireless
pressure sensor devices for the management of patients with aneurysm
repair. Unanswered
questions include potential long-term, device-related
complications, test accuracy for detecting endoleaks, the type
and number of devices that might best be used in monitoring patients,
and the extent to which the device can reduce imaging requirements following
EVAR (via direct comparison with CT). Because the impact of this
technology on clinical outcomes is not known, the TEC criteria are not
met and use of wireless pressure sensors for monitoring endovascular
aneurysm repair is considered investigational.
References
- BlueCross BlueShield Association Medical Policy
Reference Manual, Policy No. 7.01.111
- Golzarian J, Valenti D. Endoleakage after
endovascular treatment of abdominal aortic aneurysms: diagnosis,
significance and treatment. Eur Radiol 2006;
16(12):1849-57
- www.fda.gov/cdrh/pdf6/K061046.pdf
- Sonesson B, Dias N, Malina M et al. Intra-aneurysm
pressure measurement in successfully excluded abdominal
aortic aneurysm after endovascular repair. J
Vasc Surg 2003; 37:733-8
- Dias NV, Ivancev K, Malina M et al. Intra-aneurysm
sac pressure measurements after endovascular aneurysm
repair” differences between shrinking, unchanged,
and expanding aneurysms with and without endoleaks. J
Vasc Surg 2004; 39:1229-35
- Ohki T, Ouriel K, Silveira PG et al. Initial
results of wireless pressure sensing for endovascular
aneurysm repair: the APEX Trial--Acute Pressure Measurement
to Confirm Aneurysm Sac EXclusion. J Vasc Surg 2007;
45(2):236-42
- Ellozy SH, Carroccio A, Lookstein RA et al. Abdominal
aortic aneurysm sac shrinkage after endovascular
aneurysm repair: correlation with chronic sac pressure
measurement. J Vasc Surg 2006; 43(1):2-7
- Silveira PG, Miller CW, Mendes RF et al. Correlation
between intrasac pressure measurements of a pressure
sensor and an angiographic catheter during endovascular
repair of abdominal aortic aneurysm. Clinics 2008;
63(1):59-66
- Hoppe H, Segall JA, Liem TK et al. Aortic
aneurysm sac pressure measurements after endovascular
repair using an implantable remote sensor: initial
experience and short-term follow-up. Eur
Radiol 2008; 18(5):957-65
Cross References
Endovascular
Grafts for Abdominal Aortic Aneurysms,
Regence Medical Policy Manual, Surgery Policy
No. 98
Endovascular
Stent Grafts for Thoracic Aortic Aneurysms or Dissections,
Regence Medical Policy Manual, Surgery Policy No.
124
| Codes |
Number |
Description |
| CPT |
34806 |
Transcatheter placement of wireless
physiologic sensor in aneurysmal sac during endovascular
repair, including radiological supervision and
interpretation, instrument calibration, and collection
of pressure date (List separately in addition
to code for primary procedure) (new code 1/1/08) |
| |
93982 |
Noninvasive physiologic
study of implanted wireless pressure sensor in
aneurysmal sac following endovascular repair, complete
study including recording, analysis of pressure
and waveform tracings, interpretation and report
(new code 1/1/08) |
| HCPCS |
None |
|
Surgery Section Table of Contents 

|